Free calibration tools · 02
Calibration interval calculator
Enter three or more dated results to see whether an instrument looks stable or is drifting, when the trend might cross tolerance, and whether to review the interval.
Estimated trend
—
Add at least three results to estimate the trend.
- Projected tolerance crossing
- —
- Suggested interval
- —
- Observed cadence
- —
For planning only. This estimate does not account for measurement uncertainty or replace an approved calibration interval.
FAQ
Frequently asked questions
How does the calibration interval calculator estimate drift?
It divides each signed measured deviation by its positive tolerance limit, then fits a straight line over time. A fitted change of at least ten percentage points of tolerance across the entered date range is labeled drifting.
Which readings can I compare?
Use one result per date for the same instrument, measurement point, and units. Enter deviations from the same target, not the raw reading. For example, enter +0.02 if a 10.02 reading is measured against a 10.00 target. The estimate assumes a symmetric tolerance.
How is the suggested interval chosen?
The baseline is the median gap between calibration dates. The demo suggests a shorter interval, down to one day, when a reading is at or beyond tolerance or when the trend reaches a limit within two typical intervals. It only suggests a longer interval for stable readings within half the limit and with unchanged tolerances.
Is the projected crossing date reliable enough to change my schedule?
No. This straight-line estimate ignores uncertainty, conditions, and changes in use. It does not forecast more than roughly ten years from the last reading. Treat it as a prompt for review and follow your documented process and applicable requirements.